Telecordia GR-49-CORE for outdoor network interface devices
Paragraph
|
Title
|
Description
|
Benchtop Models
|
Reach-in Models
|
Other Models
|
5.1 R5-1 [53]
|
High Temperature Storage/ Mold Stress
|
70°C
|
Any benchtop or ovens
|
Any quality temperature/humidity chambers
|
Walk-in
|
5.1 CR5-1 [55]
|
High Temperature Storage/ Mold Stress
|
80°C
|
Any benchtop or ovens
|
Any quality temperature/humidity chambers
|
Walk-in
|
5.2 R5-4 [56]
|
Temperature Cycling
|
-40 to 60°C
30 minute transition
|
|
Any quality temperature/humidity chambers or Faster Temperature Cycling Chambers
|
Walk-in
|
5.2 CR5-5 [57]
|
Temperature Cycling
|
-40 to 60°C
30 minute transition
|
|
Any quality temperature/humidity chambers or Faster Temperature Cycling Chambers
|
Walk-in
|
5.3 R5-6 [58]
|
Temperature Cycling with Humidity
|
4 to 60°C
w/ RH
|
see note below
|
Any quality temperature/humidity chambers or Faster Temperature Cycling Chambers
|
Walk-in
|
5.3 CR5-7 [59]
|
Temperature Cycling with Humidity
|
4 to 60°C
w/ RH
|
see note below
|
Any quality temperature/humidity chambers or Faster Temperature Cycling Chambers
|
Walk-in
|
Don't be fooled. Test 5.3 is harder than it sounds because it requires humidity to be controlled below 10°C, which is very hard on a chamber. Also, for test 5.3, we don't recommend our benchtop series due to the requirement for humidity to be controlled during temperature transition, starting at 4°C.